CALL;
A.WIHARTA
081294124885
085778084789
087784022201
Info Produk
OUTPUT VOLTAGE RANGE RESOLUTION TEST CURRENT ACCURACY
50V( 0~ 20% ) 0~ 20M© 0.01M© 1mA @ 50K ± ( 3% + 5)
20~ 50M© 0.1M©
100V( 0~ 20% ) 0~ 20M© 0.01M© 1mA @ 100K ± ( 3% + 5)
20~ 100M© 0.1M©
0~ 20M© 0.01M© 1mA @ 250K ± ( 3% + 5)
20~ 200M© 0.1M©
200~ 250M© 1M©
0~ 20M© 0.01M© 1mA @ 500K ± ( 3% + 5)
20~ 200M© 0.1M©
200~ 500M© 1M©
0~ 200M© 0.1M© 1mA @ 1000K ± ( 3% + 5)
200~ 1000M© 1M©
1.00~ 5.00G© 0.01G© ± ( 5% + 0.1G© )
5.00~ 10.00G© 0.01G© ± ( 10% + 0.2G© )
DC Voltage
RANGE RESOLUTION ACCURACY
200V 0.1V ± ( 0.5% + 5)
1000V 1V ± ( 0.5% + 5)
AC Voltage
RANGE RESOLUTION ACCURACY
200V 0.1V ± ( 1.5% + 5)
750V 1V ± ( 1.5% + 5)
Resistance
RANGE RESOLUTION ACCURACY
20© 0.01© ± ( 1% + 5)
200© 0.1© ± ( 1% + 5)
Insulation Test Voltages : 50, 100, 250, 500, 1000V
Insulation Source Voltages : + 20% , -0%
Insulation Short Circuits Test Current : 1, 8mA nominal
Insulation Auto Discharge : discharge time 20V prior to initialization of test
Insulation Maximum Capacitive load : up to 1µ F load
Storage Temperature : -40 ~ 60° C
Operating Temperature 0 ~ 40° C
Relative Humidity : 40 ~ 75% ( 40 ~ 60% when insulation test > 1G©
Size 180 x 140 x 65nm
Weight ± 900gr